CM-1000 is a fully automated analyzer intended to quantitatively analyse chemical components in human serum, plasma, urine, and cerebrospinal liquid samples, offering constant throughput of 1000 T/h.
Sample Type:
serum; plasma; urine; cerebrospinal fluidTest Time:
Constant Speed 1000T/H, up to 61 items at the same timeMethodology:
Latex particle-enhanced turbidimetry immunoassay, Immunoturbidimetry, ColorimetryDimensions:
1100 mm(L) x 930 mm (W) x 1200 mm (H)Weight:
400 KgCM-1000
Clinical Chemistry Analyzer
Intended Use
CM-1000 is a fully automated analyzer intended to quantitatively analyse chemical components in human serum, plasma, urine, and cerebrospinal liquid samples.
l Constant throughput: 1000 T/H, optional ISE module throughput with 400 T/H.
l Double-row cleaning design: 8-step washing, more thorough cleaning.
l 270 cuvettes in total, 135 cuvettes for both of the inner and outer trays.
l Liquid level detection and empty suction detection.
l 61 test items can be detected simultaneously.
Features
Specifications
Test Principle |
Latex particle-enhanced turbidimetry immunoassay, Immunoturbidimetry, Colorimetry |
Analysis Method |
Rate Method, End-point Method, Fixed-time Method |
Test Items |
Up to 61 items tested simultaneously |
Test Speed |
Constant Speed: 1000 T/H ISE module(optional): 400 T/H |
Sample Loading Method |
Sample loading by tracks |
Sample Volume |
1.50-50.0 µL (0.1 µL increment) |
Reagent Volume |
10.0-450.0 µL (0.5 µL increment) |
Reagent Tray |
63 reagent positions for outer tray 40 reagent positions for inner tray |
Reaction Volume | 100.0-360.0 µL |
Cuvettes |
Quartz or plastic cuvettes are optional; Diameter: 5 mm; 270 cuvettes |
Test Temperature |
Temperature range: 37 ± 0.2 ℃, fluctuation value ± 0.1 ℃ |
Mixing |
Variable mixing intensity, mixing by rotation of stirring rod |
Pre-heating Time |
≤ 15 min |
Water Consumption |
≤ 20 L purified water/H |
Weight | 300 kg |
Dimension |
1160 mm (L) x 790 mm (W) x 1140 mm (H) |
Spectroscopic Method |
Multi-wavelength direct detection Optical source: halogen lamp (50 W), life of 2000 hours Light splitter: grating Detector: photodiode array Wavelengths: 1 or 2 of the following 16 preset wavelengths can be selected for testing.340, 380, 405, 450, 480, 505, 520, 546, 578, 600, 630, 660, 680, 700, 750, 800 nm Measuring points: 98 (every 7.2 sec) + 4 points for water blank |
Liver Function |
TBA, AFU, ALT, AST, ALP, AMM, GGT, TP, GPDA, GLDH, ALB, TB, DB, PA, CHE, ADA, 5’-NT, LAP, MAO, CG, GR |
Renal Function |
CRE-E, Urea, UA, β2-MG, CYS-C, MALB, NAG, RBP, CSF |
Specific Protein and Rheumatism |
ASO, RF, CRP, fCRP, Fer, CCP, Hs-CRP, TRF, RBP, C3, C4, IgA, IgM, IgG |
Glycometabolism |
GLU-HK, GLU-OX, FMN, D3-H, HbA1c, GA, LAC, NEFA |
Blood Lipids and Cardiovascular |
TG, TCH, HDL-C, LDL-C, ApoA1, ApoB, ApoE, Lp(a), HCY, NEFA, ACE, Myo, Lp-PLA2 |
Pancreatitis |
LPS, AMY |
Venous Thromboembolism and Coagulation |
D-Dimer |
Trace Elements and Electrolytes |
Ca, Mg, P, CO2, Fe, Na, Cl, K, UIBC |
Myocardial Enzyme |
CK, CK-MB, CK-MB mass, LDH, HBDH, LDH1 |
Tumor |
SA, AFU, Fer, TRF |
Gastric Function |
PG Ⅰ, PG Ⅱ |
Related Tags :
Open WeChat and Scan the QR Code. Stay Tuned with Us.